5.1 General advice
5.2 Operating conditions
®
Typical set-up: Microsart
e.jet (166MP-4)
connected to a 3-branch manifold
(16824-CS)
4 Transport and storage
– When packing the pump, make sure that it is properly secured inside the package.
– The package must be stable enough to withstand any transport damages.
Use original packaging
5 Setting up
As you set up the pump please pay attention to the following points.
– Ensure that the pump is on a stable surface and is standing upright on its
4 rubber feet.
– The pump can be mounted from underneath with four screws.
– All special safety instructions for the liquid | gas being handled must be observed.
– The head materials, hose connectors and hoses must be resistant against the
liquids being pumped.
u Check that the connections between hoses and hose barbs are connected securely.
It is sensible to secure the hoses on to the hose barbs with a clip | clamp.
This is even more important when pumping aggressive liquids or when operating
at high pressures.
– Permissible ambient temperature +5 to +40°C
– The presence of solids in the medium being handled may lead to faulty operation
of the pump.
– Permissible temperature of the medium being handled is +5 to +80°C.
– The pump should not operate against a closed system.
– The pump is limited against which pressure | vacuum it can start running.
The ideal conditions are when the pump can start at atmospheric conditions as
then the priming is guaranteed.
Transport and storage | Setting up
5