Series SAH22A SMART Air Hood
Specifications - Installation and Operating Instructions
The SERIES SAH22A SMART Air Hood
and easy to operate air flow hood on the market. By using the included hood stand
and wireless communications to a handheld device with the Dwyer Instruments
PredictAir
application software, a single operator can balance a branch in less time
™
than traditional balancing teams. Besides being lighter than most traditional capture
hoods, the ergonomic design makes the Series SAH22A easy to maneuver, with
less physical stress. The rugged polypropylene base hood features patented Quad
Flow Design Technology for controlling air flow and minimizing back pressure, which
yields superior measurement accuracy. The Wi-Fi direct communication gives reliable
communication with a distance of up to 200 yards between the hood and the handheld
test instrument. The PredictAir
application software reduces the number of steps in
™
the air flow balancing process using Predictive Balancing's Express Balance mode.
Predictive Balancing is a method of predicting the optimal flow set point for each
register and the order in which they should be adjusted.
INCLUDED WITH THE SAH22A
1. Translucent gray SAH22A base unit with 2´ x 2´ (0.6 m x 0.6 m) opening
2. Extendable pole 2´ to 4´ (0.6 m to 1.2 m) with handheld test instrument connection
3. Installation and operating manual
4. Pole safety clamp
5. Stationary Pole Adapter
6. SAH22A travel case
7. Cable adapter to connect the SAH22A
8. Charger and cable for SAH22A and handheld test instrument
MODEL CHART
Model
Description
SAH-22-IN
SMART Air Hood
balancing instrument, 4 ft (1.2 m) extendable pole;
™
batteries and handheld test instrument not included (international
shipments only)
SAH-22-LB
SMART Air Hood
balancing instrument, 4 ft (1.2 m) and 12 ft (3.6 m)
™
extendable poles; batteries not included
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
balancing instrument is the most accurate
™
Balancing Instrument
™
24.074
13.662
SPECIFICATIONS
VOLUME FLOW
Service: Air.
Volume Flow Units: CFM, l/s, m
3
/h.
Volume Flow Ranges: Supply: 40 to 2000 CFM. (68 to 3398 m3/h); Exhaust: 80 to
2000 CFM.
Accuracy at Calibration > 40 CFM: ±3% of reading ±7 CFM*.
Accuracy for Any Diffuser > 40 CFM: ±3% of reading ±10 CFM**.
Resolution: 1 CFM (1.7 m³/h) (.5 l/s).
TEMPERATURE
Units: °C, °F, °K.
Operating Range: 40 to 140°F (4.4° to 60°C).
Storage Range: -4 to 122° F (-20° C to 50° C).
Accuracy: ±0.3% of reading (no calibration required).
RELATIVE HUMIDITY
Range: 5 to 95%.
Accuracy: ±5% of reading (no calibration required).
ABSOLUTE PRESSURE
Units: mbar, Pa.
Range: 10 to 2000 mbar.
Accuracy @ 25°C and within 300 to 1100 mbar: ±2% of reading (no calibration
required).
Power Requirements: 3.6 VDC NCR18650B MH12210 lithium ion battery, not
included, user replaceable or (4) 1.5 VDC AA alkaline batteries, not included, user
replaceable.
Housing Material: Polypropylene.
Weight: 5.75 lb (2.6 kg).
Agency Approvals: CE, FCC, IC.
*Based on calibration with Sensing Precision Calibration Fixture and Dwyer
standard diffuser.
**Based on any diffuser in the Dwyer downloadable library.
Please see Bulletin TE-SAH-SW for PredictAir
instructions.