PATIENT START GUIDE
Right Leg Pictured
1
CALF CUFF APPLICATION
Wrap the cuff around the calf and secure the Velcro to hold it in place.
Make sure the wrap is snug, but not too tight.
2
Right Leg Left Leg
When both wraps are secured on
your legs, they should look like the
picture above.
4
PATIENT DEVICE USE
Unit will inflate and deflate to the
specified mode as directed by your
physician.
See accompanying documents / User' s Manual.
CAUTION: Federal Law restricts this device
to sale by or on the order of a physician.
3
TURNING THE DEVICE ON
When the wraps are secured on your
legs PRESS the RED power button for
three seconds until the light is
illuminated on each unit.
5
SWITCH MODES
For instructions on how to switch
between modes, please refer to
"Instructions" section on this page.
CONTENTS:
• 2 PlasmaFlow Devices
• Instructions for Use
The purpose of the PlasmaFlow is to aid in the prevention of Deep Vein Thrombosis (DVT) by helping to stimulate
blood flow in the legs. This is accomplished by an electronically controlled pump delivering a set amount of air
to the leg cuffs that, in turn, compress the calf or calves to aid blood flow out of the lower extremities. The
pump will inflate each leg cuff to a preset pressure of 55mmHg and deflate once the pressure is reached.
INSTRUCTIONS
POWER ON:
Press the Power Button for three seconds. The unit powers up with BLUE LED
illuminated (RED can be illuminated if battery voltage is low). The unit will be in the first working mode.
After a delay, the pumps will allow inflation of the attached wraps to a pre-determined pressure of 55
mmHg. Once the pressure reaches the proper level, the pumps will turn OFF for a 50 second "rest" period.
To enter the second mode, tap the power button to enter the second mode. Mode 2 is a step up
technology and provides a 10 mmHg incremental pressure increase at 10mmHg per step.
POWER OFF:
Push and hold the Power Button for three seconds, it will turn off.
SWITCH MODES:
In order to operate the PlasmaFlow unit in "Mode 2", simply tap the Power Button
once while the unit is powered on. The screen on the left side of the Power Button will display "0", and the
screen on the right side of the Power Button will display "F2". The unit will start operating in "Mode 2" after
a 10-second pause. To switch the PlasmaFlow unit back to "Mode 1", simply tap the Power Button once.
Mode 1: Slow Inflation:
Pressure will inflate to 55 mmHg and deflate.
Mode 2: "Step up Technology":
pause at every increment. Once the unit reaches 55 mmHg, it will deflate in the same descending increments.
BATTERY INDICATORS
In order to properly indicate the state of the battery and charger, there are TWO stages of the
BATTERY INDICATOR;
STAGE 1 – BLUE:
When unit power is ON and fully operational. Also indicates full charge.
STAGE 2 – RED:
When the battery voltage becomes low during the pumping time and rest period.
When the light is red, the battery charger must be connected immediately to avoid any interruption.
ALARMS
PlasmaFlow Alarms:
If battery voltage drops below critical level, cycling stops, audible alarm
goes on and the PlasmaFlow unit will FLASH RED. Alarm will continue for 10 seconds (unless unit is
powered OFF) and automatically turn off. Plug into power supply immediately.
"Battery Critical" Alarm
:
alarm will sound and UNIT WILL FLASH RED. Alarm will continue for 30 seconds (unless unit is powered
OFF) and automatically turn the unit OFF. Plug in power supply immediately.
Low Pressure or Leak Alarm:
pressure after 30 seconds. Make sure the wrap is snug around the calf. Turn the PlasmaFlow unit off, and
then power the PlasmaFlow unit back on. If the unit continues to alarm and the default pressure is not
reached please call ManaMed customer service at 888-508-0712. DO NOT try to open or fix the device.
Alarm Reset:
To reset an alarm condition after the "operation inhibit" stage is reached, the unit
must be turned OFF. If not manually turned OFF within 30 seconds of such an alarm condition occurring,
unit automatically turns itself OFF. Once the unit is turned back on the alarm is RESET.
DIGITAL NUMERIC DISPLAYS
Timer Indicator:
When the PlasmaFlow unit is in use, the screen on the left side of the Power
Button operates as a Timer. The screen will display digit "1" for one hour of working time, digit "2" for two
hours of working time, etc. When the Timer reaches number "99", it will automatically restart the count from
"0" (zero). When PlasmaFlow is not in use, the unit saves the previously accumulated working time (up to 99
hours). When the power is back on, the Timer will continue the count from the previously saved number.
Pressure Indicator:
When the PlasmaFlow unit is in use, the screen on the right side of the
Power Button operates as a Pressure Indicator. As the unit inflates, the numerical digits increase to display
the pressure of the PlasmaFlow unit. As the unit deflates, the numerical digits decrease.
• Patient Start Guide
• AC Adapter
The PlasmaFlow unit's pressure will increase at 10 mmHg with a
If battery voltage drops below critical level, cycling stops, audible
An audible alarm sounds when the unit does not reach desired
#QS0001, Rev. 2
™
• Charger