Descargar Imprimir esta página

Seko TEKNA EVO TPR Serie Manual De Instalación página 9

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 23
Paragraph 12 - Setting the Pause
Programming
PROG
Configuration
+
Paus
-
N.Open
pH Calibration Menu
Pressing the CAL key for 3 seconds takes you into the calibration menu. If calibration was excluded during
programming, the following appears on the display:
Calibration
Off
If calibration is active:
Cal
Calibration
Automatic
It is possible to select automatic or manual mode. In both cases, it is automatically calibrated to pH 7.
Automatic calibration:
-
The buffer solution value appears on the display. Enter the probe in the bottle and press the
second countdown necessary to complete calibration will appear on the display. If the alignment quality is below
50%, an error message appears on the display and you should press
automatically after 4 seconds). If the quality is above 50%, the value is shown on the display and, after pressing the
key, the buffer solution at pH 4 or 9 will be requested. At this point the procedure is the same as above.
-
Manual calibration:
when the buffer solution value appears on the display, insert the probe in the bottle and press the
second countdown necessary to complete calibration will appear on the display. If the alignment quality is below
50%, an error message appears on the display and you should press
automatically after 4 seconds). If the quality is above 50%, the value is shown on the display and, after pressing the
key, the value of pH 7.00 flashes on the display. Use the
your possession, then press
EM00136063
-
+
-
Calibration
Automatic Cal.
Automatic
Buffer Sol. 7 pH
-
Calibration
Manual Cal.
Buffer Sol. 7 pH
Manual
to confirm and start the calibration procedure as before.
rev. 2.2 EN
The pump can be paused by remote input.
The factory setting is Normally Open.
Changes can be made by pressing the
using the
keys to set the new value ( N. OPEN or
N. CLOSED).
Press
to confirm and return to the main menu.
Wait....
Buffer Sol. 7 pH
Quality
100%
Automatic Cal.
||||||
ok
7 pH
Error
Automatic Cal.
Buffer Sol.4/9 pH
Wait....
60s
Buffer Sol. 4 pH
Quality
100%
Automatic Cal.
||||||
ok
4 pH
Error
Wait....
Buffer Sol. 7 pH
Quality
100%
Automatic Cal.
||||||
ok
7 pH
Error
-
Manual Cal.
Buffer Sol.7.00pH
keys to enter the value of the solution in
Operation
60s
7 pH
4 pH
60s
7 pH
Wait....
60s
+
Buffer Sol.4.20pH
Quality
100%
Automatic Cal.
||||||
ok 4.20pH
Error
to exit calibration (the pump exits
to exit calibration (the pump exits
key, then
4.20pH
key. A 60
key. A 60
8

Publicidad

loading

Este manual también es adecuado para:

Tekna evo tpr 800Tekna evo tpr 600Tekna evo tpr 603