Descargar Imprimir esta página

Instruction Sheet - 3B SCIENTIFIC PHYSICS U11360 Instrucciones De Uso

Sensor de campo magnético, tangencial

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 9
3B SCIENTIFIC ® PHYSICS
Magnetic Field Sensor, Tangential U11360

Instruction Sheet

01/08 Hh
1. Safety instructions
To avoid permanent damage to the Hall sensor
in the sensor probe tip, do not subject it to any
mechanical pressure!
Do not bend the sensor probe tip!
Only use the magnetic field sensor for educa-
tional purposes!
It is not suitable for safety-related applications!
2. Description
Sensor box with attached sensor probe, 140 mm
long, 8 mm wide and 1.2 mm thick, for the meas-
urement of magnetic flux density B.
At the tip of the probe there is a Hall sensor with
an active area of about 0.5 mm², which produces a
steady output signal that is ratiometric (i.e., pro-
portional to the operating voltage).
Three buttons allow one to choose between meas-
urement ranges of 2 mT, 20 mT or 100 mT, with a
locking facility. The current setting of the meas-
urement range is shown visually by a light-emitting
diode on the left of the relevant button.
The stand clamping rod can be adjusted to hold
the sensor in the desired position and orientation
in the magnetic field that is being measured.
3. Equipment supplied
1 sensor box with permanently attached probe
1 miniDIN 8-pin connecting cable, 60 cm
1 clamping rod, length 120 mm
1 instruction sheet for U11360
4. Technical data
Measurement ranges:
Sensor type:
Sensitivity:
Nonlinearity:
Bandwidth:
1
2 mT, 20 mT, 100 mT
Linear Hall-effect sensor
Typically 1.3 mV / 100 µT
Max. ± 2.5% of the total
measurement range
1.5 kHz

Publicidad

loading