MANUFACTURER'S DECLARATION
The model 1133 Digital Blood Pressure Monitor is intended for use in the
electromagnetic environment specified below. The customer or the user of
the model 1133 Digital Blood Pressure Monitor should assure that is used
in such an environment.
Immunity
test
Electrostatic
discharge
(ESD)IEC000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
<5% U
(>95% dip in UT)
short
for 0.5 cycle
interruptions and
voltage
40% U
variations on
(60% dip in UT)
for 5 cycles
power supply
input linesIEC
70% U
61000-4-11
(30% dip in UT)
for 25 cycles
<5% U
(>95% dip in UT)
for 5 sec
Power
frequency
magnetic field
IEC 61000-4-8
Compliance
IEC 60601
level
test level
±
6 kV contact
±6kV contact
±8kV air
±
8 kV air
±
±
2 kV for
2 kV for
power supply
power supply
lines
lines
±
±
1 kV
1 kV
differential
differential
mode
mode
<5% U
T
(>95% dip in UT)
for 0.5 cycle
40% U
T
(60% dip in UT)
for 5 cycles
70% U
T
(30% dip in UT)
for 25 cycles
<5% U
T
(>95% dip in UT)
for 5 sec
3 A/m
environment-guidance
Floor should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
T
of a typical commercial or hospital
environment. If the user of the 1133
Digital Blood Pressure Monitor
Equipment requires continued
T
operation during power mains
interruptions, it is recommended that
the 1133 Digital Blood Pressure
T
Monitor Equipment be powered from
an uninterruptible power supply or a
battery.
T
Power frequency magnetic fields
should be at levels characteristic of
3 A/m
a typical location in a typical
commercial or hospital environment.
20