Recalling Measurements In Memory; Description Of Display Symbols - MABIS Healthcare SmartRead Plus Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Recalling Measurements in Memory

You can recall up to 60 total measurements (30
per memory bank) to record on a Blood
1
Pressure Log or to share with your physician or
trained health care professional.
1. Press the POWER button to turn the unit on.
Wait for the display, Fig. 1.
2. Press the
Fig. 1
memory bank 1 or 2.
3. When the correct memory bank is displayed,
press the RCAL button to display your last
stored measurement, Fig. 2.
2
4. Continue to press the RCAL button to view
the next previously stored measurement,
Fig. 3.
Fig. 2
All stored measurements will be erased by
removing and reinstalling one of the batteries.
Fig. 3
All manuals and user guides at all-guides.com
button to alternate between
16

Description of Display Symbols

SYMBOL
CONDITION
Appears when pulse
reading is displayed.
1 2
Memory banks.
1 2
1 2
Appears when battery
voltage is excessively low.
Appears when an accurate
reading cannot be
obtained.
Appears when cuff is over
pressurized.
Appears when there is too
much movement during
measurement.
Appears when unit has
malfunctioned.
17
ACTION
Record pulse reading.
Select bank for memory
storage or recall.
Replace both batteries
with new ones.
Wait a minimum of 15
minutes and remeasure.
Loosen cuff to allow blood
circulation in the wrist and
hand while waiting.
Wait a minimum of 15
minutes and remeasure.
Loosen cuff to allow blood
circulation in the wrist and
hand while waiting.
Do not talk or move during
measurement. Review
'Applying Your Blood
Pressure Cuff' section.
Unit needs to be sent
in for repair or
replacement.