HIGH LEVEL DISINFECTION
If considered clinically necessary for the patients' disease state, the device may be cleaned after each use
and disinfected once each treatment day. STERILE water must be used for rinsing.
A. BOILING
Place the four disassembled parts in a pot when boiling point has been reached and allow to boil for 5 minutes. Remove
from water and allow parts to cool and air dry thoroughly before reassembly.
NOTE: Place a small rack, on the bottom of the pot to prevent plastic parts from coming in contact with the hot bottom.
B. MICROWAVE STEAM BAG
The four disassembled parts may be disinfected in the microwave
in a steam cleaning bag (i.e. Quick-Clean
following the manufacturer's instructions. Remove from bag and
allow parts to cool and air dry thoroughly before reassembly.
Position parts in bag as shown in Figure 4.
C. ISOPROPYL ALCOHOL (70%)
The four disassembled parts may be soaked in 70% isopropyl
alcohol for 5 minutes. Rinse well and allow to air dry thoroughly
before reassembly.
D. HYDROGEN PEROXIDE (3%)
The four disassembled parts may be soaked in 3% hydrogen peroxide for 30 minutes. Rinse well and allow to air dry
thoroughly before reassembly.
E. BLEACH (Not recommended for high level disinfection)
The four disassembled parts may be soaked in 1:50 bleach to water solution (approximately 1 tablespoon bleach in
3¼ cups water) for 3 minutes. Rinse well and allow to air dry thoroughly before reassembling. You may substitute a
respiratory equipment disinfectant (such as Milton
the manufacturer's instructions for use.
CAUTION: Failure to follow the cleaning or disinfection instructions may have an adverse effect on
device performance.
DISPOSAL
The Aerobika
®
Oscillating PEP device can be disposed of with domestic waste unless this is prohibited by the disposal
regulations prevailing in your state.
NOTES
• SINGLE PATIENT USE.
• Product should be replaced after 12 months of use or immediately if damaged.
• This product has not been made (or manufactured) with BPA (Bisphenol A) or natural rubber latex.
• Do not share this medical device.
• If you are unable to exhale through the device, check the resistance setting or try cleaning the device thoroughly.
• Store the Aerobika
®
• Storage and operating range 15° C - 45° C (59° F - 113° F) at ambient to 85% relative humidity.
6
EN
†
MicroSteam
Oscillating PEP device in a dry, dust-free place avoiding continuous exposure to sunlight.