Alarm 2 (Critical)
Program Alarm 2 to
Program
desired critical set point,
Temperature
which will trigger when
the chamber warms
up and reaches this
temperature.
Alarm 2 (Critical)
The alarm can be
Set On/Off
enabled or disabled.
System Alarms
All alarms can be
Set On/Off
simultaneously enabled
or disabled.
Standby
Power saving mode.
Program
LCD will remain lit and
Temperature
temperature visible,
while the chamber
continues to warm up
until it reaches -20°C, at
which point the unit will
power off.
ID
3 digit alphanumeric
Assign system ID
system ID. Assign
each CryoPod a unique
system ID code to track
units and downloadable
temperature log data.
Low Mem
CryoPod
Set On/Off
store up to 100 hours of
log data. Low memory
indicates there is less
than 6 hours of record-
able space remaining.
Existing log data must
be downloaded to free
up space.
Time
The current time can
be programmed. This
setting is critical if you
will be utilizing the
temperature log data.
Date
The current date can
be programmed. This
setting is critical if you
will be utilizing the
temperature log data.
Once all parameters are programmed, hold SET button (B) down for 3 seconds to save
parameter settings.
All manuals and user guides at all-guides.com
Adjust tem-
perature by
1°C increments
using DOWN/UP
buttons.
Default: -140°C
Toggle between
Off and On
using DOWN/UP
buttons.
Default: On
Toggle between
Off and On
using DOWN/UP
buttons.
Default: On (Bell
on LCD)
Adjust
temperature by
1°C increments
using Down/Up
buttons.
Default: -120°C
Adjust each digit
by using DOWN/
UP buttons. Tog-
gle between each
digit by using the
SET button.
Carrier can
Toggle between
Off and On
using DOWN/UP
buttons.
Default: On
Adjust each digit
by using DOWN/
UP buttons. Tog-
gle between each
digit by using the
SET button.
Format: HH:MM
AM/PM
Adjust each
digit by using
DOWN/UP
buttons. Toggle
between each
digit by using
the SET button.
Format:
MM:DD:YY
DOWN (C)
Triple beep
UP (D)
sequence
every 10
seconds.
Pressing the
UP button
can silence
alarm.
DOWN (C)
Alarms will
UP (D)
sound for
20 minutes
before stop-
ping.
DOWN (C)
N/A
UP (D)
DOWN (C)
N/A
UP (D)
SET (B)
N/A
DOWN (C)
UP (D)
DOWN (C)
N/A
UP (D)
SET (B)
N/A
DOWN (C)
UP (D)
SET (B)
N/A
DOWN (C)
UP (D)
Add LN2 immedi-
ately, or transfer
samples to avoid
compromise if no
LN2 available. Alarm
will automatically
reset when unit
temperature is at
least 5°C colder than
critical set point.
Note LCD display
will have red back-
light, and blinking
"Alarm" indicator if
triggered.
Note LCD display
will have red back-
light, and blinking
"Alarm" indicator if
triggered.
Temperature
log data can be
downloaded onto a
PC via the included
USB cable and
external CryoPod
Data Log software.
N/A
N/A
I
English
3