Removing the data interface
– It is absolutely necessary to first disconnect the power cable from the data interface housing.
– Carefully swing housing straight back and down.
– Disconnect the ribbon cable connector (7).
– Unhook the tabs (6) from the balance rear wall.
– Cover the ribbon cable socket on the balance rear wall with the cover.
Changing the power-line voltage setting
Note: The power-line voltage must be set on the housing of the data interface.
The voltage selector switch can be set to the proper range by using a
screwdriver:
Switch position
115 V
220 V
Installing the data interface housing
– Disconnect the power cable from the balance.
– Insert the tabs (6) on the housing into the openings on the rear of the
balance.
– Plug in the ribbon cable connector (7). The loop of the ribbon cable
must lead in from the bottom.
– Swing housing up against the balance until the power connector (8) is
plugged in completely.
– Reconnect source by plugging the power cable back into connector (9).
Leerseite
Connecting to the IEEE488 bus
The 013 Data Interface Option conforms to the IEEE488 standard. The balance
can be addressed from a control instrument (Controller) as a "Talker" and a
"Listener".
– The instrument address or the "Talker Only" function is defined with bit
switches (11) (binary 0...30 and TON).
TON (Talker Only): the balance can only transmit (Talker). Addressing
– Connect the IEEE488 cable to socket (12).
– If data transfer is to be triggered by hand ("Send on Transfer" mode),
connect the handkey (42500) or foot pedal (46278) to sockets (10).
Voltage range
92...132 V
184...265 V
Microfuse: 100 mA slow-blowing
cannot be carried out.
Device Address 15