Technical description
The heating pad is equipped with an automatic
shut-off function. This switches off the heat supply
after approx. 45 min. of continuous operation.
The heating pad is fitted with a safety thermostat
that regulates the heat and prevents overheating of
the heating pad by automatically switching off the
heating pad as soon as the maximum temperature
of 55 °C is reached.
Using the heating pad
!
CAUTION! Risk of injury
• Before each use, test the temperature
with the back of your hand.
• Prolonged use at the highest level may
cause burns. Set the heating pad to the
lowest setting (Low) for continuous
use.
• If you feel any pain, switch off the
device immediately.
➤Proceed as shown to use the heating pad
(fig. 3, page 3 to fig. 4, page 4).
Cleaning and maintenance
!
WARNING! Electrocution hazard
After cleaning, wait until the heating pad
is completely dry before plugging in and
connecting to the power supply. Do not
use or store when damp!
A
NOTICE! Damage hazard
• Allow the heating pad to cool down
completely before cleaning or storing.
• Do not wash the heating pad in the
washing machine or under running
water.
• Do not tumble dry.
• Do not use sharp or hard objects or
cleaning agents for cleaning as these
may damage the product.
➤Examine the device frequently for signs of wear
or damage. If there are such signs, if the device
has been misused or does not work, return it to
the supplier before switching it on again
➤Clean the heating pad regularly and as soon as it
becomes dirty with a dry soft brush or a damp
cloth.
Replacing the fuse
➤To replace the fuse proceed as shown (fig. 5,
page 4).
Troubleshooting
Problem
The heating
pad does not
work.
The power but-
ton does not
illuminate.
The heating
pad does not
work.
The power but-
ton illuminates.
The heating
pad switches
off.
4445103576
Possible
Solution
reason
➤ Check that the
The heating
plug is fully
pad does not
inserted into
receive any
the power
power.
source.
➤ Check that the
adapter is fully
connected.
➤ Plug in another
device to
check the cor-
rect operation
of the power
source.
The heating
Check that the
pad does not
voltage supply
receive enough
provides a suffi-
current.
cient current (see
chapter "Technical
data" on page 8).
The automatic
Switch on the heat-
shut-off func-
ing pad again.
tion has
responded.
Risk of over-
Allow the heating
heating! The
pad to cool down
safety thermo-
completely before
stat has
switching on
responded.
again.
7