IMPORTANT
• Was the sample free of clear or watery fluid?
• Was the drop large enough (at least 0.3 µL)?
• Did the reaction site fill with enough blood all at once?
Step 5:
The meter will count down, showing a flashing dash, and
display result in 7 seconds.
CONSIDER YOUR RESULT. REPEAT ALTERNATE SITE
TEST:
• IF the blood sample appeared to be diluted with clear fluid.
• IF you did not vigorously rub the test site.
• IF the blood drop was not large enough to fill the
reaction site.
• IF the test was accidentally marked as a "Control" result.
• IF your result was not consistent with how you feel.
• I F more than 20 seconds elapsed from sample collection to
measurement (evaporation of the blood sample may cause
a test result that is higher than the actual value).
Any of the above situations can lead to an inaccurate test result.
All manuals and user guides at all-guides.com
IF NOT,
REPEAT TEST.
Filled
Filled
Not Filled
41