Changing the password
• Press T from the "Code" menu to enter the password setting page
• Press the "up" ( ) or "down" ( ) keys to select the first digit (flashing)
• Press "pK" to go to the next digit
• Press "T" to save the password and return to the main programming menu
Note: if you forget the password entered, the default password ("0000")
can be reset by resetting the instrument (this may also occur accidentally
if the power is cut off)
Relay programming
■ From the relay 1 (or 2) menu in normal programming, press "T" to gain access
to the relay 1 (2) programming menu
■ The parameters that can be set by the user are the following, in order:
• Operating value
• Latching
• Alarm mode
• Relay operating mode
• Set-point value
• Differential value
• Relay activation delay
■ Press "T" to go from one menu to the next
Press T from the "Delay" menu to return to the relay 1 (or 2) menu
VOLTAGE
1
➧
➧
pK
➧
1
POWER
ACTIVE
- 36 -
User Manual ADR-out Network Analyser
T
➧
pK
ACTIVE
2
POWER
T