Conversion to the
standard hydrogen
electrode potential
T 814241
04/2014
 Make sure the immersion depth is correct. The junction must be complete-
ly submersed in the solution. The junction is in the area of the bottom end
of the shaft (see arrow).
The level of the reference electrolyte must be at least 2 cm above the level
of the solution.
U
= U
+ U
H
Meas
Ref
where:
U
= ORP voltage, relative to the
H
U
= measured ORP voltage
Meas
U
= voltage of the reference system relative to the
Ref
U
is temperature dependent and can be taken from the following table
Ref
(also refer to DIN 38404-6):
T (°C)
U
[mV]
Ref
TruLine ORP 50
0
+224
5
+221
10
+217
15
+214
20
+211
25
+207
30
+203
standard hydrogen electrode
standard hydrogen electrode
T (°C)
35
40
45
50
55
60
Commissioning, measuring, testing
U
[mV]
Ref
TruLine ORP 50
+200
+196
+192
+188
+184
+180