5.input/output configuration
Switch off device
• Take the front face off.
• Select input and output calibres using the straps
as indicated on the right.
• Connect the instrument to a < 0.3 accuracy
rating multimeter
Switch on device
Input x V / x A
• Generate 0 V / A
• Set output down scale, using potentiometer P1
• Generate full scale on the input
• Set output full scale using potentiometer P2
• Adjust down and full scale settings if necessary
Note :
For any configurations not mentionned above,
please consult with us for a feasability
study.
View of board with
front face taken off
Configuration
OUTPUT
Down scale
setting
Full scale
setting
Configuration
INPUT
p5
A -AC current input configuration
B -AC voltage input configuration
C -Output configuration
Your instrument is now ready to work.
p6