Complete Dental Light; Electrical Requirements - DKL CHAIRS L2-S300 Manual De Instrucciones

Tabla de contenido

Publicidad

3.2

COMPLETE DENTAL LIGHT

3.2.1 Electrical requirements

The requirements for correct installation for any application (combined, wall, floor or ceiling)
are as follows:
Power supply
Power cable
Complete
light
version
17-24 Vac
50/60 Hz
Complete
2 x 0.5 mm2
light
300 V
version
105°C
PVC insulation
22-33Vdc
diameter
insulation 1.85
mm
Only use certified
terminals and
connectors with
resistance to
flame V1 or
similar.
Complete
light version
with
Theia
Tech
24Vac
50/60Hz
Type of power supply and
safety requirements
transformer complies with
IEC/EN 60601-1 third edition
and IEC/EN 60601-1-2 with
thermal protection or protection
downstream with at least one
appropriate fuse:
• T1.6AL 250V
Minimum requirements:
• Output: 17-24 Vac;
• Power: 26 VA;
• Class B;
• Rigidity over 4000 V.
• Thermal protection
Power supply unit complies
with IEC/EN 60601-1 third
edition and IEC/EN 60601-1-
2 with thermal protection or
protection downstream with at
least one appropriate fuse:
• T630mAL 250V
Minimum requirements:
• Output: 22-33 Vdc;
• Power: 14 VA;
• Class B;
• Rigidity over 4000 V.
Continuous protection from
short circuit or overcurrent
transformer complies with
IEC/EN 60601-1 third edition
and IEC/EN 60601-1-2 with
thermal protection or protection
downstream with at least one
appropriate fuse:
• T2AL 250V
Minimum requirements:
• Output: 24Vac;
• Power: 40VA
• Class B;
• Rigidity over 4000 V.
Thermal protection
Compliance with
Classification
IEC 60601-1
The medical
system must
be declared
compliant with
IEC/EN60601-1
by the installation
technician or
manufacturer.
Component
Note for the
built-in
installation
technician: ensure
the combined
version on
which the light
is installed is
certified to host the
complete light.
45