Solutions to some possible issues you may encounter are described in this section.
OxyGo Oxygen Concentrator
Problem
Possible Cause
Any problem
Refer to Chapter 4
accompanied
by information
on concentrator
display, indicator
lights and/or
audible signals
Concentrator does
Battery is discharged or
not power on when
no battery is present
On/Off button is
pressed
AC Power supply is not
connected properly
DC Cable is not connected
properly
Malfunction

Troubleshooting

Recommended Solution
Refer to Chapter 4
Use external power supply or
replace battery with one that is
fully charged
Check power supply connection and
verify green light is solid
Check DC Cable connection at
the Concentrator and at cigarette
lighter or auxiliary DC power source
Contact your equipment provider