SCOPE OF THIS INSTRUCTION
MASS-VIEW
®
instruments are accurate instruments for measuring and controlling the mass flow rate of gases.
These smart instruments offer multiple process values as input- or output parameters. Many parameters can be
changed on the instrument through the menu, other parameters can only be changed using our digital interfaces.
This instruction covers the quick installation for MASS-VIEW
Zeroing and Operating in just 10 steps. More detailed information can be found in documents listed below.
MASS-VIEW
®
instruments have modular instruction manuals consisting of:
Description:
Instruction manual MASS-VIEW
®
series
Operation instructions digital instruments
RS232 Interface with FLOW-BUS protocol
Modbus interface
FlowPlot Manual
Hookup diagram MASS-VIEW
®
series
Dimensional drawing MASS-VIEW
®
series
These documents can be downloaded from our website: www.massflow-online.com
CAUTION: Check MASS-VIEW
®
1
Before installing your instrument it is important to read
the label on the back of the instrument and check:
- Model number
- Pressure
- Temperature
- Power supply signal
- Flow direction
CAUTION: Check test pressure
2
Check the label on the back of the instrument and make
sure the test pressure is in accordance with
normal safety factors for your application.
CAUTION: Check if system piping is clean
3
For reliable measurement always make sure the fluid
stream is clean. Use filters to assure a moisture and
oil-free gas stream. Make sure filter surface is large
enough to avoid too high pressure drop. Install the
MASS-VIEW
®
according to the flow arrow on the
label at the back of the instrument.
2
®
instruments regarding: Start-up, Mounting,
Document number:
9.17.051
9.17.023
9.17.027
9.17.035
9.17.030
9.16.085
7.05.697
properties
CAUTION: Install instrument properly
4
Install the MASS-VIEW
®
instrument in the line and
tighten the fittings according to the instructions of
the supplier of the fittings.
CAUTION: Leak check
5
Check the system for leaks before applying (fluid)
pressure.
CAUTION: Electrical connections
6
Must be made with a standard cable or according
to the MASS-VIEW
®
hook-up diagram. (document
nr. 9.16.085)
SETUP screen
7
When the MASS-VIEW
is powered up for the first
®
time a setup menu appears where the pre-installed
gas can be selected, which flow range is to be used
and which communication type.
Warm-up time
8
Let the MASS-VIEW
warm-up for 30 minutes for
®
best accuracy.
Password
9
After entering the default password (which is 'abc') in
the security menu, all menus can be enabled or disa-
bled. Also a new password can be entered and saved.
Zeroing
10
Perform zero action if needed. See instructions at
zeroing procedure given in the Instruction manual.
(document nr. 9.17.051)
Your MASS-VIEW
®
instrument is now ready for
operation.