NOMENCLATURE

Tropic Heater Heating Units by Sammons Preston
WHITE LAMP (where applicable), when illuminated, indicates that main power is connected,
the switch is on and unit is operating.
GREEN LAMP (220/240 volt units), when illuminated, indicates that main power is connected,
the switch is on and unit is operating.
RED LAMP (where applicable), when illuminated, warns of an overheat/low water condition.
Action is necessary to correct the condition.
White Lamp
IMPORTANT: Some models are fitted with a thermal cut-out thermostat which acts as a
"non self-resetting" thermal cut-out. In the unlikely event of a fault condition, this device is
designed to open the power circuit between 190° F ± 14° F (88° C ± 8° C). It is not possible
to alter its operating range. The operation of the unit may be restored by depressing the
red button after the water/element temperatures have been allowed to cool. If the fault
condition persists, please contact Sammons Preston or an authorized agent.