Further information

Maintenance and care

NOTICE
Risk of injuries caused by electric shock!
If cleaning is required, disconnect the charger
X
from the power supply.
Clean the charger with a soft, dry tissue.
X
Never use running water or immerse the charger in water.
X
Do not sterilize or disinfect the charger.
X
After use and before placing in the charger, wipe the
X
charging pins on the hearing aids with a soft, dry tissue to
remove grease or moisture.
30

Troubleshooting

Problem and possible solutions
LED light is off:
Check if the power supply is connected correctly.
LED light is blinking red 3 times: An error in the
charger has occurred.
Take the hearing aids out of the charger and
disconnect the charger from the power supply.
Reconnect the charger to the power supply and
re-insert the hearing aids in the charging slots.
Charger has reached the temperature limit and
needs to cool down. Charging is resumed when the
temperature decreases below 45°C (113°F).