Idiomas disponibles

6.5 Insert the calibration cable into the sub-miniature female connector on the side panel.
Be sure to check that it is inserted with the proper polarity and make sure the proper
type is selected in the display.
6.6 Note : For greater accuracy, especially if there is a difference between the internal
temperature of the calibrator and the device under calibration, we recommend that
you use a thermocouple cable rather than the copper cable that is supplied with the
calibrator. The thermocouple cable must match the thermocouple type of the device
under calibration.
6.7 Connect the other end of the cable to the device to be calibrated.
6.8 Allow the proper warm up time for the Calibrator and the Device to be calibrated. At
least 30 minutes is typically recommended.
6.9 Adjust to the desired output temperature using the up/down "COURSE" adjustment
buttons on the faceplate and the "FINE" adjustment knob on the side panel. The
output temperature will be displayed on the LED display. The "COURSE" adjustment
has 16 increments and will display "high" or "low" when out of range.
7. DIRECT MV CALIBRATION AND MEASUREMENT
For directly calibration or measuring thermocouple devices in mV rather than in
temperature, press the "TYPE" button until the F or C in the LED display disappears (only
O
the "
" remains). The LED display will now indicate in mV rather than temperature units.

8. BATTERY OPERATION

WARNING: DO NOT CONNECT THE AC ADAPTOR TO THE METER IF A 9V
ALKALINE BATTERY IS INSTALLED.
8.1 For field applications, it is recommended that the operator use the rechargeable
NiCad battery that comes with
8.2 A quality 9V alkaline battery may be used in place of the rechargeable (8.4V) NiCad
battery that comes with the unit.
8.3 A flashing "F" or "C" on the LED display indicates a low battery condition.
9. CALIBRATION / REPAIR SERVICES
Extech offers complete repair and calibration services for all of the products we sell.
For periodic calibration, NIST certification or repair of any Extech product, call customer
service for details on services available. Extech recommends that calibration be
performed on an annual basis to insure calibration integrity.
10. WARRANTY
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and
workmanship for one year from date of shipment (a six month limited warranty applies on sensors and
cables). If it should become necessary to return the instrument for service during or beyond the warranty
period, contact the Customer Service Department at (781) 890-7440 for authorization. A Return
Authorization (RA) number must be issued before any product is returned to Extech. The sender is
responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit.
This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring,
operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech
specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not
be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to
repair or replacement of the product.
The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or
implied.
Copyright © 1999 Extech Instruments Corporation. All rights reserved including the
right of reproduction in whole or in part in any form.
All manuals and user guides at all-guides.com
3
433201
Ver. 1.04
11/99

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

433202

Tabla de contenido