Cleaning the Analyzer

Spills
Observe universal precautions when cleaning spills on the analyzer. Use a
10% bleach solution (1 part bleach plus 9 parts water) to clean spills, fol-
lowing the standard cleaning guidelines in "Cleaning the Analyzer" on page
9-1 of the operator's manual.
Display
Clean the analyzer's screen periodically using a soft, lint-free cloth damp-
ened with a glass-cleaning fluid or window cleaner.
The screen can be disinfected using a 10% bleach solution: apply the solu-
tion to a lint-free cloth, then wipe the screen.
CAUTION :
Do not use any cleaner containing alcohol.
Do not spray cleaner directly onto the
display — dampen the cloth instead
Air Filter
The air filter in the back of the analyzer should be cleaned at least twice
per year. Check the air filter more frequently if the analyzer is located in an
environment with excessive dust or dirt.
To clean the air filter:
Unplug the analyzer and remove the power cord from the back of
1
the analyzer.
To reach the filter, remove the air filter cover at the back of the analyzer
2
by pulling.
Wash the filter in warm soapy water and dry it completely before
3
reassembling.
Place the clean, dry air filter into its cover. Afterwards push the 4
4
clips back to their position on the analyzer.
Plug the power cord back into the analyzer.
5