Power
Display mode
Measurement mode
Measurement range
Accuracy
Normal working condition
Storage & transportation condition
Measurement perimeter of the upper arm 22cm - 42cm
Weight
External dimensions
Attachment
Mode of operation
Degree of protection
Protection against ingress of water
Software Version
1. This equipment needs to be installed and put into service in accordance with the
information provided in the user manual;
2. Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this
equipment and should be kept at least a distance d=3,3m away from the equipment.
3. (Note: As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical cell
phone with a maximum output power of 2 W yields d=3, 3m at an IMMUNITY LEVEL
of 3V/m)
All manuals and user guides at all-guides.com
Specifications
6VDC 4 x AAA Batteries / AC Adapter powered
mode: 6V
the recommended AC adapter model.
Digital LCD V.A. 84mm×73mm
Oscillographic testing mode
Rated cuff pressure:
0mmHg~300mmHg(0kPa ~ 40kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Pressure:
5°C-40°C within±0.4kPa (3mmHg)
pulse value: ±5%
Temperature: 5°C to 40°C
Relative humidity ≤85%RH
Atmospheric pressure: 86kPa to 106kPa
Temperature: -20°C-60°C
Relative Humidity 10%RH-93%RH
Atmospheric Pressure: 50-106 kPa
Approx. 368g (Excluding the dry cells)
Approx. 107mm × 103mm × 118mm
4 × AAA Batteries, Case, User Manual
Continuous operation
Type BF applied part
IP21
V01
EMC Guidance
1A (not included). Please only use
11