Emergency opening of centrifuge lid

To open the lid of centrifuge in case of power break-
down or lid lock damage, disconnect the device from
mains power supply, wait until the rotor comes to a
standstill, then lift the centrifuge and insert the in-

Rotor assembly and usage

Assembly: Place the rotor on the axis . On top of the rotor
place a special spacer - note that the pin of the spacer
must align with the groove of the axis, then place the nut
on the axis - and tighten it with the included rotor key .
Before each start, check that the rotor is firmly tightened.
Disassembly is performed in reverse sequence .
Usage: Always load the rotor symmetrically . By minimiz-
ing the difference in weight between the filled test tubes,
you reduce the wear on mechanical parts of the centri-
fuge . If the overall imbalance of test tubes in the rotor ex-
USER MANUAL
Centrifuge CM-7S Plus
cluded emergency lid opening key in to the groove
and move the lever to the side (as shown in the fig-
ure below) . The lever is located on the bottom of
the centrifuge, in the groove near the control panel .
EMERGENCY LID OPENING LEVER
ceeds 7 grams, the centrifuge will start emergency brak-
ing it reaches 1000 rpm and the display will show when
imballance error .
ATTENTION! Do not use damaged rotors!
ATTENTION! Samples may contain pathological ma-
terial, including pathogens of serious diseases. Always
check test tubes for damage before centrifugation.