Signs And Symbols; General Notes; Emc Information; Maintenance And Repairs - NATURE & DECOUVERTES 15198170 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Immunity Test standard
IEC 60601-1-2 Test Level
Electrostatic discharge (ESD)
IEC 61000-4-2
+ 2 kV, + 4 kV, + 8 kV, + 15 kV air
+ 2 kV, for power supply lines
Electrical fast transient/burst
+ 1 kV signal input/ output
IEC 61000-4-4
100 kHz repetition frequency
+ 0,5 kV, + 1 kV differential lines
Surge IEC 61000-4-5
+ 0,5 kV, + 1 kV, + 1 kV common mode
0 % U
 ; 0,5 cycle. At 0°, 45°, 90°, 135°,
T
Voltage dips, interruptions and variations
180°, 225°, 270° and 315°.
on power supply input lines
0 % U
 ; 1 cycle and 70 % U
T
IEC 61000-4-11
Power frequency (50/60 Hz) magnetic
field IEC 61000-4-8
Conducted RF IEC 61000-4-6
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
Note: U
is the a.c. mians voltage prior to application of the test level
T

SIGNS AND SYMBOLS

The following symbols appear in this manual, on the therapy lamp or on its accessories:
Specifies serial number of the
therapy lamp
Daylight ON/OFF
Authorized Representative in
the European Community
Ingress Protection Rating
DC polarity
Compliance Level
+ 8 kV contact
+ 8 kV contact
+ 2 kV, + 4 kV, + 8 kV, + 15 kV air
+ 2 kV, for power supply lines
N/A
100 kHz repetition frequency
+ 0,5 kV, + 1 kV differential lines
N/A
0% U
 ; 0,5 cycle. At 0°, 45°, 90°, 135°,
T
180°, 225°, 270° and 315°.
 ; 25/30 cycles;
0% U
; 1 cycle and 70 % U
T
T
Single phase: à 0°.
Single phase: at 0°.
0 % UT ; 250/300 cycle
0% U
; 250/300 cycle
T
30 A/m
30 A/m
50 Hz/60 Hz
50 Hz/60 Hz
3 V
0,15 MHz – 80 MHz
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
80% AM at 1 kHz
Read operating instructions
before use
Caution: Consult accompanying
documents
Class II Equipment
Date of Manufacture
Manufacturer
14

GENERAL NOTES

Important
- This appliance is not intended for commercial or
clinical use, but only for individual, private house-
hold use.
- If you are taking medication such as pain relievers,
antihypertensives or antidepressants, consult
your doctor before using this appliance.
- People with retinal diseases as well as diabetics
; 25/30 cycles;
T
need to be examined by an ophthalmologist be-
fore using this appliance.

EMC INFORMATION

3 V
Warning
- Use of this appliance adjacent to or stacked with other device should be avoided because it could
result in improper operation. If such use is necessary, this appliance and the other device should
be observed to verify that they are operating normally. If the appliance does not light on or flickering,
it means that the appliance is not operating normally.
- Don't near active HF surgical equipment and the RF shielded room of an ME system for magnetic
resonance imaging, where the intensity of EM disturbances is high.
- Do not change any cable without authorization of the manufacturer. Use of cables other than
those specified or provided by the manufacturer of this appliance could result in increased elec-
tromagnetic emissions or decreased electromagnetic immunity of this appliance and result in im-
proper operation.
- Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm to any part of this appliance, including power
cable. Otherwise, degradation of the performance of this equipment could result.
- If the appliance does not light on or flickering due to the electromagnetic disturbances, do not use
it and try to place the appliance in other location.
- For maintaining the basic safety and essential performance with regard to electromagnetic disturbances
for the expected service life, please precaution to take if the use location is near (e.g. less than 1,5 km
from) from AM, FM or TV broadcast antennas.

MAINTENANCE AND REPAIRS

- Avoid exposing the screen to direct sunlight, spotlight or any other heat source.
- Do not use alcohol or ammonia base liquid to clean the screen. The cleaning information provided
in the section "Cleaning and Storage".
- Please do not attempt to repair the appliance yourself. This could result in serious injuries. Failure
to observe this regulation shall void the warranty.
- If this appliance is damaged in any way or it is not working, it should not be operated. Please try
to fix the problem according to FAQ table. If the problem is still not resolved, please contact cus-
tomer service or the dealer, since special tools are needed to make repairs. Never try to repair the
appliance yourself.
- Please do not use if you have an eye disease such
as cataracts, glaucoma, diseases of the optic
nerve in general and inflammation of the vitreous
humour.
- If you have health concerns of any kind, contact
your general practitioner!
15

Publicidad

Tabla de contenido
loading

Tabla de contenido