Battery replacement

Prior to battery replacement, disconnect the instrument from any connected test
leads. Only use batteries as described in the technical data section!
Switch off instrument. Disconnect test leads.
Loosen the screws of the battery cover on the instrument rear. Lift the battery
cover.
Remove discharged batteries.
Insert new batteries 1.5 V IEC LR03.
Replace the battery cover and retighten the screws.
Please consider your environment when you dispose of your one-way batteries or
accumulators. They belong in a rubbish dump for hazardous waste. In most cases, the
batteries can be returned to their point of sale.
Please, comply with the respective valid regulation regarding the return, recycling and
disposal of used batteries and accumulators.
If an instrument is not used over an extended time period, the accumulators or batter-
ies must be removed. Should the instrument be contaminated by leaking battery cells,
the instrument has to be returned for cleaning and inspection to the factory.

Fuse replacement

Prior to fuse replacement, ensure that multimeter is disconnected from external
voltage supply and the other connected instruments (such as UUT, control
instruments, etc.)
Only use fuses as described in the technical data section! Using auxiliary fuses,
in particular short-circuiting fuse holders is prohibited and can cause instrument
destruction or serious bodily injury of operator.
Switch off the instrument. Disconnect test leads.
Loosen the screws on the instrument rear.
Lift the case cover.
Remove the defect fuse.
Insert new fuse.
Replace the case cover and retighten the screws.
Fuse (A): F 400 mA / 600 V Ceramic 6.3 x 32 mm
Fuse (A): F 10 A / 600 V Ceramic 6.3 x 32 mm