2.0 Specifications
Table I:
Type of measurement
Accuracy of measurements
Power requirements
Operating flow range
Operating temperature range
Available communication
Warranty
Calibration
Certification/Compliance
3.0 Installing the batteries
This unit uses two AA alkaline batteries.
polarity (⊕ and ⊖) correctly aligned. Close the cover. (Figures 2 and 3.)
To install batteries, extend the wire stand. Open the cover. Insert the batteries with the
• Load the new batteries with their polarity (⊕ and ⊖) aligned correctly.
Precautions for battery replacement:
•
Do not use rechargeable batteries.
Figure 2: Extend the wire stand before
opening battery door.
Volumetric flow
Accuracy of ± 2.00% of flow reading or ± 0.200 mL/min, whichever is greater.
2 AA alkaline batteries 1.5 V DC each/3 V DC 200 ma
0.50 to 500 mL/min
32 °-120 °F (0 °- 48 °C)
USB type B data port
One year (excludes recalibration)
NIST traceable
Yearly recalibration is recommended
CE, Ex, Canadian ICES-003, WEEE, RoHS 2, China RoHS 2, UKCA (see section 10.0)
– +
Figure 3
Insert the batteries
as marked.
www.restek.com
–
+
5