16) Error Messages
Code / Fusion report
ERROR 2.0 !
(MAIN_FREQUENCE_LO_ERROR)
report sentence: "FRQLEr"
ERROR 2.1 !
(MAIN_FREQUENCE_HI_ERROR)
report sentence: "FRQHEr"
ERROR 3 (NO_FREQUENCE)
Report name "FREQErr"
ERROR 4 (NO_VOLTAGE_IN)
Report name "LinErr"
ERROR 4.0 !
(MAIN_VOLTAGE_LO_ERROR)
report sentence: "LinLEr"
ERROR 4.1 !
(MAIN_VOLTAGE_HI_ERROR)
report sentence: "LinHEr"
ERROR 20 (HIGH_TEMP_ERROR)
Report name "HTAErr"
ERROR 21 (LOW_TEMP_ERROR)
Report name "LTAErr"
ERROR 23
(TRASFO_TEMP_ERROR)
Report name "THTErr"
ERROR 24 !
(CHSS_TEMP_HIGH)
report sentence: " ChsHTE "
ERROR 30 !
(LOAD_OPEN_ERROR)
report sentence: " NLoadE "
ERROR 32 (NO_LOAD_ERROR)
Report name "NLoadE"
ERROR 33 !
(LOAD_MEASURE_ERROR2)
report sentence: "LoadE2"
ERROR 34.0
(OUT_CURRENT_ERROR)
report sentence: "IoutEr"
ERROR 34.1!
(CURRENT_SHC_ERROR)
report sentence: "I2SCHE"
ERROR 34.2 !
(OUT_SHC_CURR_ERROR)
report sentence: "I1SCHE"
ERROR 35 (V_OUT_ERROR)
Report name "VoutEr"
ERROR 36 !
(LOAD_MEASURE_ERROR3)
report sentence: " LoadE3"
Cause
Network frequency too low
Network frequency too high
Error that identifies irregularity or failure in the
network frequency.
Error that identifies the device switching off for
lack of power supply.
Main voltage too low
Main voltage too high
Error that identifies the ambient temperature is
too high.
Error that identifies the ambient temperature is
too low.
Error that identifies the processor temperature is
too high. The internal protection disables the
device.
Frame temperature too high.
Electric circuit open in the fitting. The error is
displayed only during fusion cycle.
May be subsequent to a false contact or coil's
break
Error that indicates an open circuit.
It is only displayed during welding.
This error is typical for open circuit or interruption
of a wire during welding.
Resistance in the fitting too high
Current output not correct
Before the fusion:
- current output request higher than possible.
E.g.: electronic limitation for safety reasons
During the fusion:
- current output request higher than possible.
Error that identifies anomaly in the achievement of
the device output voltage.
- Possible short circuit or overload.
- The fuse of the power supply is damaged.
A difference between coupling resistive
parameters and those indicated in the barcode
have been found. The fitting may have
Pag. 24/50
Solution
Check power generator: refer to chapter 18 for power
generator specification.
Change energy supply. Connect the control unit to
national grid.
Check power generator: refer to chapter 18 for power
generator specification.
Change energy supply. Connect the control unit to
national grid.
Check the power supply quality when connected to a
generator.
Check the network cable quality or integrity, or the
power source where the device is connected, or
verify whether the power source has the necessary
features to service the needs of power.
Check power generator: refer to chapter 17 for power
generator specification.
Change energy supply. Connect the control unit to
national grid.
Check power generator: refer to chapter 17 for power
generator specification.
Change energy supply. Connect the control unit to
national grid.
It is not possible to perform the welding because the
environmental conditions are out of the temperature
range.
It is not possible to perform the welding because the
environmental conditions are out of the temperature
range.
Leave ECU powered allowing the fan to remove the
excess of heat and cooling down.
Heavy use of the control unit. Poor ventilated
environment.
Shut off the control unit and wait until the frame has
cooled down
Check quality contact on terminal pins. Avoid
oxidation, wear or bad contact.
Change damaged or failed fitting
Check the quality of the fitting terminal contacts.
Replace the broken or defective fitting.
Check the contact between the control unit's
connectors and fitting's terminal pins. Avoid
oxidation, wear or bad contact.
Wrong data in the fitting's barcode. Change fitting
Check the contact between the control unit's
connectors and fitting's terminal pins. Avoid
oxidation, wear or bad contact.
A short circuit in the fitting could be happened.
Replace the fitting and repeat welding procedure.
Check fitting's request of current is within control unit
limits.
Check fitting for short circuit.
Replace the fitting.
Check fitting for short circuit.
In case of manual fusion data input: check fitting's
request of current is within control unit limits.
Replace the fitting and repeat welding procedure.
Replace the fitting.
Contact technical support at no. +1 770 632 7530
It is recommended to control nature and matches of
the product that you want to use. Replace the fitting.
In case of emergency try to shift from the automatic