byko-visc RT
6 Viscosity Measurement
1.
If protective cap is attached, turn it clockwise to remove.
2.
Move spindle attachment upwards, insert spindle and release attachment.
3.
Tap measurement icon in main menu to access measurement screen.
4.
Lower unit into sample and tap green Start button.
2
4
5
1
2
3
4
5
6
7
8
9
10
11
12
Touch areas of the screen with measured data to access corresponding setting.
10
1
3
6
7
8
11
9
12
TTS / TTT
Counter for Time to Stop or Time to Torque
Viscosity
Average of measurements - taken every spindle revolution, tap unit
to change
Maximum
Maximum value, depending on current spindle and speed
Speed
Current spindle speed in rotations per minute (RPM), tap value to
change
Spindle
Identifier for current spindle, tap entry to change
Torque
Current value in percent: Values < 15% or > 95% indicate that speed
or spindle settings should be changed (invalid measurement)
Temp.
Current temperature of sample in °C or °F (if temperature probe is at-
tached), tap to change unit
Graph
Live updating graph of viscosity over time
Start
Green Start button: Start and resume measurement
Pause
Blue Pause button: Pause measurement
Stop
Red Stop button: End measurement and save results
Help
White Question mark: Open context-sensitive help
BYK-Gardner GmbH
10
300 003 476 - 2301