1.3
Warnings and Cautions
Please retain these instructions for use for future reference.
• WARNING: The Sunset Heated CPAP Tube is validated for use with two specific
humidifiers: Philips Respironics DreamStation Series and Philips Respironics
System One. Do not use the Sunset Heated CPAP Tube in combination with other
Humidifiers.
• WARNING: The Sunset Heated CPAP Tube used with incompatible parts can result
in degraded performance.
• WARNING: Failure to use a mask or accessory that minimizes rebreathing of
carbon dioxide or permits spontaneous breathing can cause asphyxiation.
• WARNING: The user is accountable for the compatibility of the Sunset Heated CPAP
Tube and the humidifier and all of the parts and accessories used to connect to
the patient before use.
• WARNING: Always connect-disconnect the tube from the humidifier by using the
grips on the cuff. Do not pull directly on the tube.
• WARNING: To prevent disconnection of the tubing or tubing system during use,
only use accessories in compliance with ISO 5367 or ISO 80601-2-74.
• WARNING: Do not add any attachments to the Sunset Heated CPAP Tube,
otherwise, the humidifier and its accessories might not function correctly,
affecting the quality of the therapy or cause injury to the patient.
• WARNING: When used routinely, check the tube daily for holes, deformation, or
any other form of damage. When damaged, replace the tube immediately.
• WARNING: Operating the Sunset Heated CPAP Tube close to other heat sources
including lamps, electric blankets, etc., can cause elevated airflow temperatures
that can affect the quality of the therapy or cause injury to the patient
• WARNING: The tube shall be replaced after 90 days of use. Using the Sunset Heated
CPAP Tube after the maximum 90 days duration of use can affect the quality of
therapy or cause injury to the patient.
• WARNING: Do not sterilize the Sunset Heated CPAP Tube.
• WARNING: The Sunset Heated CPAP Tube is for single patient reuse only. Using the
Sunset Heated CPAP Tube on multiple patients can lead to contamination and injure
the patient.
• WARNING: Do not subject the tube to pressure, crushing or other mechanical stresses,
as damage to the tube can affect the quality of the therapy or cause injury to the
patient.
2