BEFORE FIRST USE
¬ Fully charge battery using the Spin
Touch USB cable and adapter plug in
AC outlet.
¬ Ensure your Spin Touch has the
latest firmware. A USB connection to
a Windows® PC is required:
1. Download and install the WaterLink
Connect 2 application for Windows
at softwarecenter.lamotte.com.
2. Plug meter into the computer with
the provided USB cable and launch
WaterLink Connect 2 application
from the Start Menu.
3. A prompt will appear if firmware
updates are available. Select
Update. Testing and data transfer
will not be possible until firmware
has been updated.
A4 QRC
See the WaterLink Spin Touch
manual for additional information.
FILLING: Single-Use Disks
When the syringe is placed in the water
sample, and the plunger is pulled all the way
up, the syringe will hold more than enough
water sample to adequately fill the disk. Hold
the syringe vertically and insert the tip into the
fill hole in the disk. Press the plunger slowly
and smoothly to fill the disk.
2. Do not overfill the disk. If the disk is
overfilled, sample water will flow out of the
overflow hole in the center of the disk. The
disk is not leaking. Dry the disk and run the
test.
4. Do not introduce air bubbles into the disk.
The reagent chambers will not fill entirely
and results will be inaccurate. As soon as
a bubble starts to form, pull back on the
plunger to draw the bubble out of the disk.
Begin the filling process again.
6. Disks should be filled and used within 10
minutes. They cannot be filled ahead of
time.
TESTING
1. Press
and hold until the meter turns on.
2. Tap
. Select a sanitizer type. Tap
3. Tap
. Select a disk series (found on disk packaging). Tap
limited by Sanitizer Type selection.
4. Remove disk from packaging.
5. Use syringe (1189) to fill disk with pool or spa water sample.
6. Insert disk. Cover disk with Universal Disk Cover (1719). Close lid.
7. Tap
to start test. Tap
8. Results will be displayed.
9. Select an option.
- Tap
to add a tag.
- Tap the highlighted
to save the test results to the test log if Auto Save is not enabled.
- Tap the hightlighted
to send the results to the enabled Mobile Bluetooth Printer.
- Tap the highlighted
to send the results to a Bluetooth enabled device.
-
Tap
to return to the Test Screen.
10. REMOVE THE DISK FROM THE CHAMBER. DISPOSE OF THE DISK. Disks left in the chamber can
leak and damage the device. Do not store the disk cover on the hub.
11. Press and hold
for 2 seconds to turn meter off.
NOTE: For water samples over 100 °F (38 °C) subtract 0.1 from pH result or, for the most accurate result, wait
until water sample is below 90 °F (32 °C) to test.
Remove salt residue daily. Salt will damage the meter and cause inaccurate results. See CLEANING section in
manual.
1. Fill the disk using slow, even
pressure. The sample water will fill
the spaces between the baffles in a
counterclockwise order. Each space will
fill from the bottom to the top. Sample
water should be added until the sample
water in the fourth chamber fills to the
top of the chamber slightly past the
embossed fill line. It is OK to fill slightly
past the fill line.
3. Do not under fill the disk. If the disk
is under filled, the reagent chambers
will not fill entirely and results will be
inaccurate.
Overfilled
(Dry SpinDisk before placing in meter)
5. Wet disks should be dried thoroughly
Air
with a lint free wipe. The disk should be
Bubble
handled by the edges.
For filling and troubleshooting tips go to:
waterlinkspintouch.com/support.html
2
QUICK
START GUIDE
to confirm.
to cancel the test. If the test is cancelled discard the disk.
English
to confirm. NOTE: Disk Series are
Correct Fill
(Fill line)
Underfilled