5.Specifications
Measuring Range
32.0℃~42.0℃
32.0℃~35.9℃
Accuracy
36.0℃~39.0℃
39.1℃~42.0℃
Resolution
0.1℃/0.1℉
Measuring Distance
5cm
Operating Temp.
10℃~50℃
≤85%
Relative Humidity
Auto Power Off
Meter shuts off automatically after 7 seconds of inactivity
Power Supply
9V battery
Weight
145g
134*88.5*36mm
Dimensions(L*W*H)
Note: the accuracy is obtained under ambient temerature18℃-28℃/64℉-82℉.
6.Meter Description
A / B. Infrared sensor
C. LDC display
D. Button (trigger) for measurement
E. Battery cover
7.Operating Instruction
7.1 Operating steps:
7.1.1 Installing the battery
7.1.2 Hold the meter by its handle grip and point it toward the parts to be measured with distance of 5cm, press the
measuring trigger, measuring finished when the meter sounds a "tick"
89.6℉~107.6℉
89.6℉~96.6℉
±0.3℃/0.5℉
96.8℉~102.2℉
±0.2℃/0.4℉
102.3℉~107.6℉
±0.3℃/0.5℉
50℉~122℉
7.1.3 Release the trigger to read the measuring data
Note: 1. If the meter has been stored in a cold or hot environment, allow it 15 to 20 minutes to acclimate to room
temperature before making measurements.
2.Make sure the skin is dry and that no hair interferes with the measurement.
8.Battery Replacement
◆ When the low battery icon"
" appears on LCD, replace the meter's battery.
◆ Open the battery compartment, replace the 9V battery and close
the battery compartment cover.
9. Maintenance
◆ Do not use volatile liquids to clean the unit, swipe it with dry soft cloth.
◆ Do not disassemble the unit, repair it by qualified personnel.
◆ Do not immerse it in water.
◆ Do not store it in high temperature or humidity.
10. Troubleshooting
10.1 LCD shows "HI":over-range measurement, that is, the measuring data is above 42.0℃/108.5℉
10.2 LCD shows "LO": under-range measurement, that is, the measuring data is below 32℃/89.6 ℉
11. Accessories
◆ User's Manual
◆ 9V Battery