TROUBLESHOOTING
If your Neb-u-Lite EV Nebulizer Compressor fails to
function, consult the following Troubleshooting guide.
If the problem persists, consult your Graham-Field
authorized distributor.
Problem
Device doesn't
operate
Nebulization is
weak
JB0112-061-INS-LAB-RevD14
Cause and Solution
Check if plug is properly fit into a grounded electrical outlet.
When device has been run continuously for more than 30 minutes
just before using, auto-shut-down may have been activated by the
built-in thermal protector; ensure that device is OFF for at least 30
minutes before next usage.
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to
compressor or nebulizer cup; replace as needed.
Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.
Check if filter is dirty; replace filter as needed.