Calcium sensor usage
The calcium sensor's electrode uses a PermaFil (non-
refillable),
ion-selective
chamber is gel filled and sealed off; therefore, no further
reference filling solution is required.
The sensor is pre-calibrated at the factory. No need for
calibrating during the first 12 months from receiving.
After 1 year and if the sensor has been well maintained
we recommend doing recalibration. For information on
where to purchase calibration solution as well as
required calibration software please reference the ion
selective product pages at www.neulog.com. If there are
any questions please contact us at
Usage guide:
Concentration range
pH range
Temperature range
Minimum sample size 3mL in a 50 mL beaker
Reproducibility
Electrode resistance 1 to 4 megaohms
Equipment required:
Wash bottle with distilled or deionized water.
Several clean beakers.
1 mL, 10 mL, and 50 mL pipettes (a 50 ml graduated
cylinder can also be used).
V2018.02
19
electrode.
-7
1M to 5x10
M
0.02 to 40,000 ppm
2.5 to 11 pH
0 to 40°C
±4%
The
reference
.
info@neulog.com