Medline Hemo-Force MDS600INT2 Instrucciones De Operación página 14

Bomba para tvp intermitente 2nda generación
Ocultar thumbs Ver también para Hemo-Force MDS600INT2:

Publicidad

Idiomas disponibles

Idiomas disponibles

2. Indicator Light
i.
Green: Power button illuminated in green to indicate the control
unit is running on AC power.
ii.
Amber: Power button illuminated in amber to indicate the control
unit is running on battery power.
iii. Flashing Amber: Flashing Alarm Mute button in amber to indicate
a fault has been detected followed by an audio alarm.
3. Audible Alarm
Press the alarm mute button and the audible alarm
can be temporary muted. The audible alarm will go off
again after 10 minutes if the error still persists.
Contraindications & Cautions
Do not use the Hemo-Force® where increased venous or
lymphatic return is not desired. The Hemo-Force® Intermittent
Compression System could be contraindicated and should not be
used by persons with known or suspected deep vein thrombosis,
pulmonary edema, congestive heart failure, severe arteriosclerosis,
thrombophlebitis, or active infection. Do not use it on persons
with painful and sensitive extremities, where the garment will
interfere with gangrenous limbs, on persons with vein ligation
or recent skin grafts, or extreme deformity of the leg.
NOTE: If you are unsure of any contraindications, please refer to the
patient's physician before using the Hemo-Force® Intermittent Pneumatic
Compression System.