• Turn on by pressing the on/off switch and start therapy;
• During the treatment session, the nebuliser cup can be rested on its own support;
• Carry out treatment preferably seated and in a comfortable position;
• When the aerosol jet becomes intermittent, interrupt the treatment for a few seconds to allow the suspended
drops of solution to fall on the nebuliser cup walls. Resume the treatment session, which will be finished when
no more nebulised solution is released from the nebuliser cup;
• Once the treatment session is completed, switch the device off by unplugging it, remove the accessory used
and clean the device and accessories according to the instructions in the Cleaning and Maintenance paragraph.
• The patient and operator are the same person. The steps that the patient/user can do safely are only those
specified in the paragraphs "Preparing and using the appliance, " "Cleaning and maintenance of the compressor,
nebuliser cup and accessories. " Warning! Do not service and/or repair the appliance while it is being used! Any
other maintenance/service activity/operation must be done solely by a PiC Service Centre.
COMPRESSOR CLEANING AND MAINTENANCE
Warning! Do not wet the device or put it into water or other liquids. For cleaning, use a clean, dry cloth only.
CLEANING AND MAINTENANCE OF THE NEBULISER CUP AND ACCESSORIES
Remove and detach all the nebuliser cup components and dip them in boiling water for about 5 minutes.
WARNING! Do not boil the masks.
Rinse the accessories (mask, mouthpiece, fork and tube) under running water. Cold disinfect all parts that come in
contact with the patient using denatured alcohol. Carefully dry all accessories and the device before reassembling.
Store the device and the accessories in a cool, dry place, away from light and heat.
Never use benzene, thinners or other inflammable chemical substances for cleaning.
For increased hygienic safety, never use the same accessories on more than one patient. Purchase a dedicated kit
for each user (Aerosol Kit
Clean and check that the condition of the accessories before and after use. Replace them if damaged.
WARNING! Never leave the nose fork unattended as it is a small part that, if swallowed by children,
could represent an ingestion and suffocating hazard
CHECKING AND REPLACING FILTERS
Periodically check the condition of the filter. The filter was inserted to protect the compressor.
Correct filter maintenance prolongs the life of the aerosol. The frequency with which filter should be replaced
depends on the conditions in which the device is used. The filter is located in the upper part of the aerosol.
TO REPLACE THE FILTER:
• Remove the filter cover by lifting it up carefully, using a tool with a flat blade;
• Remove the filter to be replaced from its seat;
• Carefully fit the new filter;
• Reposition the filter cover by pushing it down fully in its seat, making sure that it cannot be removed without
the use of a tool.
WARNING! Never leave the removed filter cover and/or filters unattended: These are small parts posing
the risk of suffocation if swallowed by children.