3. SPECIFICATIONS
3.1 ELECTRICAL
Ranges: 240 V, 2 kV, 6 kV, 11 kV, 22 kV, 33 kV, 132 kV, 275 kV
Detection Frequency: (40 to 70) Hz
Power Supply: (3) 1.5 V C cell batteries
Power Consumption: 40 mA
Signal Lamp: 47 lux @ 20 cm
Sound Volume: 75 db @ 1 m from detector
Electrical Safety: Compliant with EN 61326-1, EN 61000-4-2, EN 61000-4-3,
EN 55011
3.2 MECHANICAL
Case Height: 9 in (229 mm)
Case Width: 3.78 in (96 mm)
Case Protection: IP65 per IEC 526
Weight: 1.30 lbs (0.59 kg) batteries included
Shock Test: IEC 68-2-29
Vibration Test: EN 61010, clause 8.3
Drop Test: EN 61010, clause 8.4
Impact Test: EN 61010, clause 8.2
3.3 ENVIRONMENTAL
Operating Temperature: (5 to 130) °F (-15 to 55) °C
Storage Temperature: (-4 to 150) °F (-20 to 65) °C
Humidity: (0 to 93) % RH @ 104 °F (40 °C)
Cold Temperature: IEC 68-2-1
Dry Heat: IEC 68-2-2
Damp Heat: IEC 68-2-3
14
Non-Contact, High Voltage Detector Model 275HVD - User Manua