Ammonium sensor usage
The ammonium sensor's electrode uses a PermaFil (non-
refillable),
ion-selective
electrode.
The
reference
chamber is filled with gel and sealed; therefore, no
reference filling solution is required.
Usage guide:
-6
1M to 5x10
M
Concentration range
0.01 to 18,000 ppm
pH range
4 to 10 pH
Temperature range
0 to 50°C
Minimum sample size 3mL in a 50 mL beaker
Reproducibility
±4%
Electrode resistance
1 to 4 megaohms
Equipment required:
Wash bottle with distilled or deionized water.
Several clean beakers.
1 mL, 10 mL and 50 mL pipettes (a 50 ml graduated
cylinder can also be used).