Tank Capacity Mode Training
1. While holding
show on the display. This indicates that the Tank Capacity Mode is currently off.
2. Turn the Tank Capacity Mode on by pressing
Alternate Numeric Display and Percent Indicator will turn on.
3. Press
to proceed to the first step of training. 0% will flash in the Alternate Numeric
Display indicating that the weight of an empty container must be registered.
4. Apply an empty container (or a load representing the weight of an empty container) on
the platform. Wait 5 seconds for the load to stabilize, then press
the Alternate Numeric Display indicating that the weight of a full container must be
registered.
5. Apply a full container (or a load representing the weight of a full container) on the
platform. Wait 5 seconds for the load to stabilize, then press
training process, and the SET TANK indicator will turn off.
6. From this point forward, the Alternate Numeric Display will show the percent of material
remaining in the container from 0% to 100%.
NOTE: The 100% weight must be larger than the 0% weight by at least 200g. If not, "Err"
will show on the Alternate Numeric Display. To clear the error, retrain as above using a
100% weight that is at least 200g larger than the 0% weight. During training, the ZERO key
may be used to assist in setting the 0% and 100% values. This will have no effect on the
training, as the Tank Capacity Mode relies on absolute, not relative, weights.
Perform the following steps to turn the Tank Capacity Mode Off:
down, press
. The SET TANK indicator will turn on, and OFF will
OCTA-WIRELESS User's Guide
. The display will show ON, and the
. This completes the
. 100% will flash in
11