improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they
are operating normally.
The use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the FIT Pro, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
WARNING: If the use location is near (e.g., less than 1.5 km from) AM, FM or TV broadcast antennas, before using this
equipment, it should be observed to verify that it is operating normally to assure that the equipment remains safe with
regard to electromagnetic disturbances throughout the expected service life. When the AC input voltage is interrupted, the
equipment will stop battery charging and if the power supply restored, it could be recovered automatically. This degradation
could be accepted because it will not lead to unacceptable risks and it will not result in the loss of basic safety or essential
performance.
12.3 Product Disposal
This product contains electrical and electronic equipment.
1. The internal components of this product may contain hazardous materials. Dispose of this product according to
local or regional waste administration systems and regulations. Caution: Hazardous materials present in electric
and electronic equipment may harm human health and the environment if the equipment is not properly
disposed of.
2.
Do not dispose of electric or electronic equipment together with unsorted municipal waste.
3.
For further information regarding the environmental performance of this product please visit our website
www.hygeiahealth.com. The breast pump and parts are made of plastics which have an impact on the
environment when disposed of. Please recycle according to local regulations.
13. OPERATIONAL & STORAGE CONDITIONS
Operational Conditions
14. TRAVEL
!
WARNING: Use ONLY the AC power adapter recommended for the Hygeia FIT Pro Breast Pump.
INFORMATION: This breast pump contains a lithium-ion battery. Carrying such batteries on airplanes may be restricted.
Confirm airline policies prior to travel. Please consult with the country you are visiting to find out if there are any restrictions
that pertain to traveling with lithium-ion batteries.
ART-70295-01-AA
-4
-20
Transport/Storage Conditions
ENGLISH | 29