Move the cursor to Calibration CL and press
Enter.
Move the cursor to Gain and press Enter
Adjust (if necessary) the value of reading by
pressing
to the value of DPD1 reading
test and then confirm by Enter.
POTENTIOSTATIC CHLORINE PROBE CALIBRATION
Attention, it will be necessary to do a comparative free chlorine test (for example a DPD1 photometric test)
ZERO CALIBRATION is executed at factory and then go directly to gain calibration.
V-MS is been projected so that the electrical zero probe calibration is equal to electrochemical zero calibration.
So if it's really necessary to calibrate the zero of the probe, read the following instructions:
Switch off V-MS and open the connection box.
Disconnect the chlorine probe input connection from the chlorine board (refer to the next figure).
Switch on V-MS and wait until the display reaches a stable reading (any value).
Rev 6.0
C
a
→
>
C
a
O
→
>
G
G
a
i
R
e
a
→→→ Disconnect this 4 poles connector
Disconnect all cables connected to this connector. Make a
short circuit with a piece of cable between pin 1 and pin 2 and
reconnect the connector (leave empty pins 3 and 4)
Cod. 00.008.764