Dip Switch Set-Up
To set-up the KT6 for your specific application, you
might need to change some of the Dip Switches.
The Dip Switches are located underneath the chart.
Use a pen or small screw driver to flip the switches.
Remember to install correct chart to match cor-
responding Dip Switch setting.
Slide toggle to this side for on
O
1
All switches are
N
shown in off
position.
Slide toggle to this side for off
Dip Switch #1 is not active on KT6
Dip Switch #2 is not active on KT6
Recording Time:
KT6 has two different recording time options:
1 day
7 day
°F/°C
You can record in °F or °C with the KT6 by using
Dip Switch #4.
°C
°F
Temperature Range
KT6 Dip Switches for ranges
0 to 100°
0 to 50°
Calibration
Your instrument was carefully tested and calibrated
before being shipped from the factory. additional
calibration is not required. However, should cali-
bration be desired in the future, we recommend
that it be at our lab, Call Customer Service at
630-543-3747. If you wish to calibrate yourself,
follow these procedures:
1.
Identify a standard that is more accurate than
this unit.
2.
Place the probe(s) of the unit and the standard
into a controlled environment, allowing unit
to completely stabilize. For best results,
calibrate the unit at levels typically monitored
during normal operation.
3.
To activate the calibration mode, turn the unit
off. now press the ON/OFF key and the
ALARM key at the same time. The unit is in
the "User Calibration" mode when the red
LED is on. The display will show a C during
the warm up period and then will flash
between 0 and F or 0 and C (depending on
2
3
4
5
6
7
Figure 5
#3 On
#3 Off
#4 On
#4 Off
#5 Off
#6 Off
#7 Off
#5 On
#6 Off
#7 Off
50 to 100
0 to 250
8
0 to 500°F (only F)
-50 to 50
0 to -50
0 to -100
Alarm Speaker:
The alarm speaker may be activated or deactivated
with the use of dip switch #8.
Deactivated
Activated
The Alarm speaker will sound during minimum,
maximum or probe missing alarm conditions.
Note: Dip switch settings are only updated when
the unit is turned on.
measurement selected.) Match the reading of
the recorder to the reading of the reference
standard.
4.
To raise the temperature press the CHART
ROTATION (Up Arrow) key. To lower the
temperature press the HOME (Down Arrow)
key.
To ensure you have matched the standard
allow the pen to stabilize for at least 30
seconds before exiting calibration mode. If
your KT6 has one pen, skip to step 6.
5.
Press the alarm button for 2 seconds to switch
from Pen 1 to Pen 2. Repeat step 4 for Pen 2
temperature calibration.
6.
When calibration is complete simply press the
ON/OFF key to save the calibration settings.
Calibration is stored in memory even after you
turn the unit off. User calibration information will
not be lost if AC power fails.
6
#5 Off
#6 On
#7 Off
#5 On
#6 On
#7 Off
#5 Off
#6 Off
#7 On
#5 On
#6 Off
#7 On
#5 Off
#6 On
#7 On
#5 On
#6 On
#7 On
#8 On
#8 Off