the top of your hand.
4. Hold the child upright and insert the aspirator tip into the nostril to seal the
nostril. SEE FIG.B
5. Take care not to insert the aspirator tip too deeply into the nostril and not to
press on the inner wall of the nostril.
6. Press and hold the start-stop button to aspirate and remove secretion.
Suctioning only works when the start/stop button is held down.
7. Suction only for a few seconds in the nostril, then repeat as necessary.
8. To avoid eardrum pressure imbalance due to vacuum, children should not
hold their breath, close their mouth or swallow while using this product.
9. Using when the container is full may result in flooding of the motor with
secretions and damage to the device. Always empty the container if you see
the waste in the container approaching the MAX line. SEE FIG.C
10. Always remove the container and clean after using the device. Putting the
appliance down with a full container may cause secretions to overflow and
be drawn into the device the next time it is started.
11. In standby mode the unit switches off automatically after 1 minute.
NOTE: Regular use of saline and/or seawater for nasal cleansing and mois-
turising may help clear the nose. Read manufacturer's instructions or consult
your doctor before using.
08. CLEANING AND MAINTENANCE
1. The tips and the secretion container must be removed from the device
and washed separately. They should be rinsed under running water with a
toothbrush and mild detergent.
2. The tips and the container can be disinfected in water at 100°C for no longer
than 10 minutes.
3. Always wash the silicone tip and container after use.
4. Make sure all parts of the device are dry before reassembling to prevent the
build-up of bacteria and flooding of the motor.
5. Do not use abrasive materials or caustic cleaners when cleaning the device.
This may adversely affect its service life and performance.
6. Switch off the device while cleaning.
7. Do not immerse the device under water.
8. Clean the aspirator base only with a clean, damp cloth.
09 SPECIFICATIONS
Input: 5V 1A | Power: 5 W | Rechargeable battery: Lithium 3.7V, 1100mAh | Max-
imum negative pressure: 58 kPa | Tip material: silicone | Noise level: ≤59 dB |
Charging time: approximately 2.5 hours | Device operating time: up to 2 hours |
Weight: 133g | Dimensions: 103x51x42 mm