Vaina para procesos asépticos

Publicidad

Idiomas disponibles

Idiomas disponibles

5. Commissioning and operation
5.2 Mounting instructions for EHEDG and 3-A
Observe the following instructions, especially for EHEDG-certified and 3-A marked
instruments.
For compliance with the EHEDG certification, one of the EHEDG-recommended
EN
process connections must be used, see marking in data sheet TW 95.61.
For compliance with 3-A conformity, a 3-A marked process connection must be used,
see marking in data sheet TW 95.61.
Mount the electrical thermometer, including the thermowell / protection tube, with
minimal dead space and able to be cleaned easily.
Mounting position of the thermowell / protection tube should be designed to be
self-draining.
The mounting position must not form a draining point or cause a basin to be formed.
5.3 Cleaning in place (CIP) cleaning process
Only use cleaning agents which are suitable for the seals used.
Cleaning agents must not be abrasive nor corrosively attack the materials of the
wetted parts.
Avoid thermal shocks or fast temperature changes. The temperature difference
between the cleaning agent and rinsing with clear water should be as low as
possible. Negative example: Cleaning at 80 °C [176 °F] and rinsing at 4 °C [39 °F]
with cold water.
For tank-mounted sensors the tank cleaning devices shall be positioned in such a
way that the sensor can be assessed and is perfectly cleaned.
12
WIKA operating instructions protection tube, model TW61