2 Installation

2
Installation

Environmental Considerations

There are several factors that may affect the performance of your STATMATIC. Please review these factors, and
select a suitable location in which to install the device.
• Temperature and Humidity
Avoid installing your STATMATIC in direct sunlight or close to a heat source (e.g. vents or radiators). The
recommended operating temperatures are 5 – 40°C (41 – 104°F) with humidity of 25 – 85%.
• Environment
The STATMATIC should be operated in a clean, dust free environment.
• Work Surface
The STATMATIC should be placed on a flat, level, water-resistant surface. Never install and operate the device on
a sloped surface.

• Electromagnetic Environment

The STATMATIC has been tested and meets applicable standards for electromagnetic emissions. While the device
does not emit any radiation, it may itself be affected by other equipment which does. We recommend that the
device be kept away from potential sources of interference.

Electrical Requirements

Use properly grounded and fused power sources with the same voltage rating as indicated on the label at the back of
your STATMATIC. Avoid multiple outlet receptacles. If using a surge suppressor power bar, plug in one STATMATIC
only. The power input plug on the back of the unit is to be used as a separating device to the power supply network.

Media Connections

Plug power plug into socket.
Fasten and secure compressed air hose onto coupling.
On the manometer the compressed air must indicate 4 – 4.5 bar (58 – 65 psi).
In case of a deviating display, pull out handle on the compressed air regulator and turn clockwise or counter-
clockwise until 4 – 4.5 bar (58 – 65 psi) are displayed.
Push handle on the compressed air regulator back in and lock it.
To ensure secure and faultless function only use compressed air according to the information contained in the
technical data sheet.

Check Attachment Handpiece Adapter to Adapter Holder

(see chapter 3 Operation)

Insert Spray Can

(see chapter 4 Maintenance)