Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 13
5. Press the rotary knob to continue with the descaling
step 2.
 The descaling step 2 on the gauge illuminates as
green.
6. Subsequently rinse the device with the rinsing set
until no more lime deposits are flushed out.
The "service" program ends here for units that have
been set to operate with deionised water.
7. Press the rotary knob to continue the descaling step
3.
 The descaling step 3 on the gauge illuminates as
green.
8. Close the ball valve.
9. Use a funnel to fill the dissolved liquid Elma Steam
Descaler into the device.
To avoid discolouration on the filler opening, wipe off
any splashes from the Elma Steam Descaler
immediately.
10.
Close the pressure tank screw.
Press the rotary knob to continue the descaling
11.
step 4.
 The descaling step 4 on the gauge flashes green. The
device heats up (warm-up time).
 The descaling step 5 on the gauge illuminates as
green (warm up time is completed).
Reaction time! Duration, approximately 2 hours.
We recommend to let the solution react overnight.
You can switch off the device during the reaction
time. The pressure will be dissipated in the pressure
vessel overnight. If the device is switched on after the
reaction time, then the gauge automatically switches
to descaling step 6 The prerequisite hereby is that the
pressure vessel is depressurised!
English - Art. No. 1130878
[5]
Feb-24