The Deuterium lamp will be switched on if a wavelength less than 325.0 nm is
selected for a Gamma. The current data will be lost if the wavelength is changed.
INTEGRATION
This sets the integration time for which the result is measured.
The current data will be lost if the integration time is changed.

DELAY TIME

This sets the time between pressing RUN and the start of the measurement.
The range is from 0 to 99 minutes and 59 seconds.
FACTOR
Only available in CONC mode, the factor is used to multiply the absorbance result
to produce a concentration result. Changing the factor will not cause current results
to be lost. They will be recalculated using the new factor.
UNITS
ENTRY
Only available in CONC mode, used to enter the required description of the
concentrations up to 11 alphanumeric characters.

AUTOPRINT

When ON, results are printed automatically after each run.
Before attempting to print at any point during operation of the instrument ensure
that the printer is ready to print, ie switched on, on line and supplied with paper.
Failure to do so will result in an error condition. Press CLEAR to clear the error
message ( the system may take a little time to respond ). Ensure the printer is
ready and retry.
CELL PROG
REF MODE
GD Issue 1 (08/2000)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
FIXED
:
Enter integration time in seconds.
:
Set a time in the range 00.05 to 99.59,
using . to separate minutes and seconds.
:
Enter the factor for the concentration.
:
Enter units for concentration using the TEXT
page.
:
Toggles between ON and OFF.
:
Switches to the CELL PROG. screen (if fitted).
:
Toggles the status of the Cell Programmer reference
mode (if fitted).