Quirumed SA252F Manual De Instrucciones página 12

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

PS: After sterilization, the balance water and pressure in the chamber will be return to the water
reservoir automatically.
4. You are now, ready to take the power cord out from autoclave, and plug the female end into the
standard grounded AC outlet.
PS: POWER SOCKET 240VAC / 15A 220VAC / 15A
CAUTION: Before plug the power cord, please turn-off the power switch.
5. Check the power lamp, if it is light on, that means the power is already stand-by now.
CHAPTER 4 OPERATION
PLEASE CHECK CHAPTER 3. AT FIRST.
1. Open the door and turn filling / draining water valve to open position, until the water level cover
the heater completely than put the sterilized instruments on the trays, sterilization box or spring
holder into the chamber.
PLEASE DON'T FORGET TO PUT THE STERILIZATION INDICATOR INTO CHAMBER.
2. Set-up the 121ºC or 134ºC by temperature control knob (under the door catcher).
3. Closed the door and turn the door handle to tight.
PS: PLEASE PUT BACK THE DOOR SAFETY CATCHER, BEFORE YOU PRESS "START " SWITCH.
4. Set up the programs as you want (Push power switch on at first).
a. Set-up the sterilization time.
b. Set-up the dry time. The program is starting working when press "START" switch.
AS TO THE TIME SET, PLEASE REFER THE ENCLOSED APPENDIX.
5. After the chamber pressure arrival you set-up. The LED of sterilization timer (See figure 1) turn to
"flash" that means the sterilization timer starts work Then the LED light on. This means the steriliza-
tion timer finished working then starts the dry timer.
6. The LED of dry timer (See figure 1) turn to "flash" that means the dry timer starts work. Then the LED
light on. This means the dry timer finished working then the buzzer start working
7. 40 seconds buzzer will report you, "MY DEAR MASTER! YOUR COMMAND HAVE BEEN COMPLETED. "
CAUTION: - BEFORE OPENING THE DOOR ENSURE THE PRESSURE GAUGE IS AT "ZERO" POSITION.
- PLEASE DON'T FORGET THE CHECK THE STERILIZATION INDICATOR, AFTER STERILIZATION CYCLE
- TO USE THE UNIT CONTINUOUSLY, BE ABSOLUTELY SURE TO OPEN THE DOOR AND ALLOW THE UNIT
TO COOL OFF FOR AT LEAST 20 MINUTES.
Items to be sterilized
All Items need put into sterilization box.
If the sterilization item put to mach then the cove of sterilization
box cannot put on.
By the way, don't forget open both side windows before put steri-
lization box into chamber.
Fit the end of the heater cover into the grooves of the chamber
partition fitting, and check that there is no play.
Please see figure 3.
12
www.quirumed.com

Publicidad

Tabla de contenido
loading

Tabla de contenido