6. Use only a dedicated power supply.
7. The silicon rubber gasket on the inner door and the front of the chamber should be kept clean.
8. The water filter cleaning gate at rear of the machine should be cleaned at least every season.
9. We recommend use of chemical indicator strips as a check for sterility. These strips may also be
kept as a record of sterilization.
10. Do not over load or overfill the water into the water reservoir.
11.
Attention! Hot surface
12.
Protective earth (ground)
13.
Off, disconnection from the mains.
14.
On, connection to the mains
15.
Consult instruction manual for use
16. Disposal of Electrical & Electronic Equipment (WEEE): This product should be handed over to an
applicable collection point for the recycling of electrical and electronic equipment.
For more detailed information about the recycling of this product, please contact your local
city office, household waste disposal service or the retail store where you purchased this
product. (European community only)
17.
Date of manufacture
It is a 6-digit number. The first 4 digits represent the year and the last 2 digits represent the month.
18.
Authorised representative in the European
EC
REP
19.
Manufacturer
CHAPTER 6 APPENDIX
6-1 PROGRAM DETAILS AS FOLLOWS:
SETTING UP THE STERILIZATION TIME:
* PRESSURE ON 2.1 kgf/cm² (APPROX. 134ºC / 274ºC)
* PRESSURE ON 1.2 kgf/cm² (APPROX. 121ºC / 250ºC)
* DRY TIME: 30 minutes
* SPECIAL SELECTION RANGE:
# FOR REFERENCE : 1 kgf/cm² = 0.98 bar = 14.2 psi.
# Sterilization instruments must be made by metal without plate, or autoclave plastic or rubber
material.
# Liquid sterilization is available, but it must adjust the venting system by our engineer. So, as for
liquid sterilization must be noticed by your purchase order.
14
www.quirumed.com
WRAPPED Set-up 15 minutes sterilization time.
UN-WRAPPED Set-up 4 minutes sterilization time.
WRAPPED Set-up 30 minutes sterilization time.
UN-WRAPPED Set-up 22 minutes sterilization time
LIQUID Set-up 40 minutes sterilization time.
Liquids sterilization need special optional device. Request by order.
TEMPERATURE SELECTION RANGE 118ºC ~134ºC
STERILIZATION TIME SELECTION RANGE 0 ~ 60 minutes
DRY TIME SELECTION RANGE 0 ~ 60 minutes