RANGE LOCK
This instrument has a Range Lock feature that displays readings to
the best resolution, shown by / on the display. This can be
turned off by pressing the
same display resolution for the entire pressure range. This is useful
to stop the display switching between resolutions where there are
fluctuations in pressure. To cancel this mode, press the
again.
ADVANCED FUNCTIONS
The following additional functions apply to models 2080P,
2081P, 2082P, 2083P, 2084P, 2085P
LOGGING
These models have functions that enable the user to store and
retrieve up to 250 readings and output such to a PC or Epson
compatible printer via Infra-red DigiLink. We also
offer DigiLog, a customised Windows™ software package.
Note: Some functions may be inhibited while information is stored in
either Auto, Manual or Max/Min modes. This is to prevent
inadvertent parameter mixing whilst in any of the above modes.
To re-activate these functions please delete any stores information
by following the "Output of data" section.
SETTING INTERNAL CLOCK
Press d
ecimal point
on display with the current year and month settings
Key in new year and month if desired and press
Display will show current date of month setting (DD). Key in new
value if desired and press
Display will show current hours and minutes setting (24 hour clock).
Key in new values if desired and press
will then return to normal mode.
LOGGING ON DEMAND
This function allows readings to be stored as and when required. A
maximum of 50 readings can be stored this way.
When ready to store a reading press
and will appear on the display. It is possible to enter up to a four
0
digit reference number including decimal point. Press
enter and store reading, time, date and reference number.
Instrument will return to standard mode, however
remain on display until data is erased.
PRESET INTERVAL LOGGING
This function allows readings to be taken automatically at a preset
time interval.
NOTE: If data is already stored, as signified by the
symbols appearing on display, this function will not operate.
Press
key.
Enter required time interval in hh:mm (ie, in order to take a reading
every 1 minute, key in 0001) and press
return to display and
will continue logging until a maximum of 250 readings are taken or
data is down-loaded.
A R
/
(•) key and the
key to accept.
NE
symbol and
symbol will remain on display.
A R
key. The display will then use the
and
NE
key to accept. Instrument
key.
0000
will appear on the display.
key. Current reading will
A R
/
symbols will appear
(YYMM).
key to accept.
,
NE
symbols
key to
symbol will
or
Instrument
key